LABOSTAR RO DI

Optionally, a charged 0.2 µm sterile filter at the dispenser can be used to remove bacteria and endotoxins. An uncharged 0.2 µm sterile filter is also available. The 7 l storage tank capacity can be increased by an additional 30 or 60l tank.

Informasi Pemesanan:

Telpon: (+62-21) 84304918 | (+62-21) 84304853
 Email: mbi@majabintang.com

Categories: , Tags: ,

EVOQUA LABOSTAR® RO & RO DI

High Quality, Low Price
The LaboStar® RO and RO/DI series produce high quality water with very low energy consumption. The use of “low energy” membranes enable a very efficient and economical operation. All LaboStar​ RO and RO/DI systems are equipped with a built-in 7 l storage tank. Pure water is collected in the tank and is recirculated by an integrated pump.

LaboStar RO produces Type III water. LaboStar RO/DI is equipped with an additional final DI cartridge. This cartridge polishes the RO water to Type II quality < 0.1 µS/cm. Both units offers two extraction ports: water can be extracted under pressure from the Point-of-use dispenser or from the outlet on the back of the unit.

Optionally, a charged 0.2 µm sterile filter at the dispenser can be used to remove bacteria and endotoxins. An uncharged 0.2 µm sterile filter is also available. The 7 l storage tank capacity can be increased by an additional 30 or 60l tank.

LaboStar RO/DI systems are delivered with the first set of all modules and filters required for immediate operation.

LABOSTAR RO/DI ADVANTAGES

  • Connection to municipal drinking water supply
  • 7 liter built-in storage tank
  • Pure water circulation right into the dispenser head
  • Easy to dispense water using the practical POU dispenser
  • 0.2 µm sterile filter with or without endotoxins retention as option available
  • Whisper operation mode
  • Rapid and simple disinfection
  • Conductivity monitoring of RO and DI water
  • Easy module exchange via quick-release connections

TYPICAL APPLICATIONS

  • IC
  • Pathology
  • Cytology + histology work
  • Buffer preparation
  • Photometry and spectrophotometry
  • Media preparation
  • Electrophoreseses
  • General Chemistry
  • Feed for laboratory ultra-pure water systems
  • Final Rinse of laboratory washing machines

SPECIFICATIONS

System Performance Unit LaboStar 10 RO ​LaboStar 10 RO DI
Delivery flow rate l/h ​10 ​10
Salt retention % 98 98
Bacteria retention % 99 99
Particle retention % 99 99
Permeate conductivity ​µS/cm based on feed water < 0.1
Bacteria* ​cfu/ml ​<1 ​<1
Particles > 0.2 µm* per ml <1 <1
Max. delivery flow rate ​l/min 1.2 1.2
Delivery flow rate @ 0,5 bar l/h 70 70
Delivery flow rate @ 1,0 bar l/h 65 65
Feed water specifications
Pressure bar 0.1 – 5 0.1 – 5
Conductivity µS/cm < 2000 < 2000
Silt density index SDI < 3 < 3
Free Chlorine ​mg/l ​< 0.5 ​​< 0.5
Total Iron mg/l < 0.1 < 0.1
CO² mg/l 15 15​
Temperature °C 5 – 35 5 – 35
Shipping weight ​kg ​23 ​24
​Power Supply v/Hz 100-240/50/60 100-240/50/60
Dimension: H x W x D mm 535 x 290 x 400 535 x 290 x 400
Item code ​W3T324492 W3T324493

* = only in combination with a 0.2 µm sterile filter

CONSUMABLES AND ACCESSORIES

Item Number
​Description ​Change Frequency
W3T197613 Replacement Pre-treatment module AMB 3 – 6 months*
W3T197620 Replacement RO Module 10 l/h 2 – 3 years
W3T197618 Replacement DI Cartridge VMD 3 – 6 months*
W3T199279 Sterile filter 0.2 µm with endotoxin retention (pack of 3 filters) 6 months
W3T199209 Sterile filter 0.2 µm (pack of 3 filters) 6 months
W3T199880 Vent Filter for built-in 7 l storage tank (pack of 3 filters) ​12 months
W3T197552 CO2 Trap CT1 with Vent Filter VT1 Kit for 30 and 60 l tank ​12 months
W3T324494 30 l Extension storage tank ​-
W3T324495 60 l Extension storage tank ​-
​​W3T199991 Wall bracket ET30 for 30 l tank ​-
W3T197563 Wall bracket ET60 for 60 l tank ​-
W3T314413 CO2 Degassing unit (max. 150 l/h) ​-

* = Cartridge exchange may be more frequent. Subject to feed water quality and consumption.